Loading...

Spectrophotometer with USB connectivity and wide wavelength range
Download CatalogSmall sample volume (only 1.0 mL required)
Power Supply 230 VAC, 50/60 Hz
Halogen lamp (long-life light source, approximately 2,000 hours)
Accurate wavelength selection (motor-controlled wavelength selection)
Two types of test tubes supported (square cells and round test tubes can be used without adapter)
PC connection and data output (USB connection enables data output via PC operation)
| Wavelength Range | 340 to 1000 nm |
| Spectral Bandwidth | 8 nm (all wavelengths) |
| Wavelength Accuracy | ±2 nm (at 585 and 879 nm)* |
| Minimum Wavelength Scale | 2 nm (equal intervals) |
| Optical System | Single beam, diffraction grating |
| Stray Light | 0.5 %T or less |
| Display | LCD |
| Measurement Range | Absorbance -0.200 to 3.000 ABS / Transmittance 0.00 to 200%T / Concentration 0.000 to 1999C |
| Photometric Accuracy | ±2 %T |
| Photodetector | High-sensitivity silicon photodiode |
| Light Source | Halogen bulb |
| Sample Volume | 1.0 mL or more |
| Measurement Tube | Square cell, round test tube |
| Power Supply | 230 VAC, 50/60 Hz |
| Dimensions | 265 (W) × 265 (D) × 165 (H) mm |
| Weight | Net 4.8 kg (main unit) |
| Square Cell | 2 units |
| Round Test Tube | 10 units |
| USB Cable | 1 cable |
| CD-ROM (Software) | 1 disc |
| AC Adapter | 1 unit |
| Vinyl Cover | 1 sheet |
| Instruction Manual | 1 manual |
*Specifications and appearance may be changed without notice for improvement.